Lumex Bariatric Easy Lift Features
- Lower base height maximizes stability and fits under most beds
- Rubber base coating protects furniture and walls
- Ergonomically designed hand pendant with pendant hook
- Warning beep for low battery charge and weight capacity levels
- Easy access emergency stop button
- Emergency manual lowering device
- White powder-coated finish
- Storage bag included for slings and charger
- 24 volt DC motor allows quiet smooth operation
- Maximum Weight Capacity 600 lb
What to buy with Lumex Bariatric Easy Patient Lift
How to use Lumex Patient Lift LF1090?
Assembling Lumex Patient Lift
- Remove the nuts and bolts from top and bottom of mast.
- With one hand on the mast and another on the boom, pull the mast to an upright position.
- Reinstall the bolts in the holes at bottom of mast and hand tighten them.
- Attach boom to mast and hand tighten bolt and nut.
- Attach the L-shaped arms to the back of the knee pad.
- Attach knee pad assembly to the mast.
- Place the control box in the bracket.
- Insert the actuator plug into the control box.
- Insert the pendant plug into the control box.
- Tighten all nuts
Charging The Batteries of Patient Lift
- Ensure the control box power switch is ON (i.e., the red button is up).
- Insert the charger plug into the battery box charging socket (the top slot).
- Insert the other charger plug into the power supply.
- All LED indicator lights should be ON during charging.
- When one green LED indicator light illuminates, the batteries require approximately two to three hours to be fully charged.
- When only the red LED indicator light illuminates, the batteries require approximately seven to nine hours to be fully charged.
- NOTICE: Do not charge the batteries for more than twelve hours.
- Unplug the charger FIRST before using the lift.
- Info: The lift will not operate while the batteries are charging.
The lift will not operate while the red reset button is down.
- Replace the batteries when frequent charging is required.
Graham Field Power Lift Operating Instructions
- Turn on the power by turning the Red Reset Button clockwise.
- Press the pendant UP or DOWN button once.
- After the battery indicator lights de-illuminate, push the pendant up button and down button to ensure the lift is operating correctly.
- Position the lift base around or under the patient.
- Widen the base.
How to position Patient on Lift Sling?
- Position the patient in a sitting position.
- Slide the sling down patient’s back to lumbar position.
- Draw the shoulder straps to the front of the patient close to the chest.
- Draw the waist belt around patient’s waist and press hook and loop sections together to fasten.
- Place shoulder straps inside the positioning loop.
- Attach the shoulder straps to the hooks.
How to lift the patient using Lumex Patient Lift System?
Lifting Patient From Bed
- Fit sling as described above.
- Push lift towards patient. Open the base of the lift. Apply the brakes on both rear casters.
- Position patient’s feet on the foot platform and knees against the knee pad.
- Attach the sling straps to the hooks.
- Have the patient’s hands holding the handles. For patients who cannot hold the handles, have them hold their arms around the chest.
- Press the “UP” button on the hand pendant.
- Before the patient’s body is lifted from the bed, stop and make sure the sling is secured and patient’s knees are against the knee pad.
- The patient should be comfortable with the pressure under the arms. If not, adjust the sling and try another loop option to release pressure.
- Press the “UP” button until the patient’s body has completely left the bed.
- Transfer to another object can be done at this position without the patient attending a full standing position.
- Release the brakes, close the base, and pull the lift away from the bed.
- Lower the patient to the object intended.
- To lift patient in a standing position, continue to lift until the patient’s knees are locked in a standing position.
- At the standing position, patient’s knees may move away from the knee pad.
- Close the base, and pull the lift away from the bed.
- Lower the patient to the object intended.
- Reverse the above procedures when lowering the patient to the bed
Lifting Patient From Chair
- Fit sling as described in “Fitting Lumex Sit-To-Stand Sling”.
- Push lift towards patient. Open the base of the lift to go around the chair. Apply the brakes on both rear casters.
- Position patient’s feet on the foot platform and knees against the knee pad.
- Attached the sling straps to the hooks.
- Have the patient’s hands holding the handles. For patients who cannot hold the handles, have them hold their arms around the chest.
- Press the “UP” button on the hand control.
- Before the patient’s body is lifted from the chair, stop and make sure the sling is secured and patient’s knees are against the knee pad.
- The patient should be comfortable with the pressure under the arms. If not, adjust the sling and try another loop option to release pressure.
- Press the “UP” button until the patient’s body has completely left the chair.
- Transfer to another object can be done at this position without the patient attending a full standing position.
- Release the brakes, close the base, and pull the lift away from the chair.
- Lower the patient to the object intended.
- To lift patient in a standing position, continue to lift until the patient’s knees are locked in a standing position.
- At the standing position, patient’s knees may move away from the knee pad.
- Release the brakes, close the base, and pull the lift away from the chair.
- Lower the patient to the object intended.
- Reverse the above procedures when lowering the patient to the chair.
For more information please refer to Lumex Easy Lift Patient Lifting System User Manual
Lumex LF1090 Patient Lift Safety Guide
WARNING: Cancer and Reproductive Harm - www.P65Warnings.ca.gov
Lumex Bariatric Easy Lift Patient Lift Specifications
Particulars
|
Dimensions
|
Floor clearance
|
5"
|
Spreader bar minimum height |
22"
|
Spreader bar maximum height |
71" |
Overall length
|
50"
|
Minimum base width
|
26"
|
Maximum base width |
43" |
Maximum weight capacity
|
600 lb
Evenly Distributed |
Product weight
|
137 lb
|
Graham-Field Lumex Patient Lift Warranty
Lift frame and spreader bar: 3 years
Actuator: 2 years
Parts, including casters, control box, pendant, charger: 2 years
Batteries: 1 year